Model 410N4X, Process Analyzer for Oxygen, Wall Mounted Corrosion-resistant NEMA4X (IP65) Enclosure Rating

Specifications

Method of Detection

Customer replaceable electrochemical 'fuel cell' O2 sensor. Expected life 3 to 5 years depending on typical measurement concentrations.

Ranges Available

Any range from 0-2.0% to 0-100.0% O2

Resolution

0.1% O2

Accuracy & Repeatability

Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale

Drift

Less than 2% of full scale per month

Response Time (T-90)

6-8 seconds to 90% step change

Ambient Temperature Range

32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package

Linearity

Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale

Size & Weight

Dimensions will vary depending on enclosure style and options required

Power

115 VAC, 60 Hz220 VAC, 50 Hz

Output Options

4-20 mA into 500 ohms non-isolated standardIsolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs

Alarms

High and/or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC ratingLow flow alarm

Analyzer Type

O2 Analyzer

Primary Gas(es) to be Measured

Oxygen (O2)

Hazardous Area Rating Options

General Purpose, Non-Hazardous

Applications

Features


  • Best value continuous O2 analyzer
  • Bright digital readout
  • Long life electrochemical O2 sensor
  • Fast response (T90 - 6 to 8 seconds)
  • Microprocessor control for push button calibration
  • Easy-to-maintain modular layout
  • Built in sample pump or pressure regulator
  • 4 - 20 mA outputs