C06689-A2C | TELEDYN Vietnam


Teledyne Oxygen Sensor A-2C ,C06689-A2C

Application overview
RANGE
BACKGROUND GAS
TELEDYNE INSTRUMENTS
SENSOR OPTIONS
Trace
Less than 1000 ppm CO2 in the background gas
311, 316, 317, 318, 3160
B-2C
3000T series
L-2C
More than 1000 ppm up to 100% CO2 in the background gas
All trace oxygen instruments
A-2C
Note: To use the A-2C or B-2C sensor in a 3000T series instrument, adapter part number B66378 is required.
Note: The L-2C sensor cannot be used in our older analog models (300 series).
Percent
Less than 1000 ppm CO2 in the background gas
311P, 320A / P / BRC, 326, 327, 328
B-1, B-3 standard;
C-3 for high concentration of O2
More than 1000 ppm up to 100% CO2 in the background gas
A-5
CLASS
(click for spec control drwg.)
PART No.
APPLICATION
COMMENTS
        A-1: Obsolete, contact factory
C6689-A2C
Used for trace O2 measurements in CO2 containing backgrounds (0.1-100% CO2). CO2 must be present.
Use where acid gases are present (CO2, HCl, HCN, etc). CO2 must be present.